LAMP Molecular Analyser
These analysers employ LAMP technology, delivering fast, reliable results that are particularly valuable for diagnosing a full panel of infectious diseases including sexually transmitted infections, respiratory diseases and tuberculosis. The single-channel analyser is versatile, suitable for onsite or self-test home use.
Our Single Channel LAMP Molecular Analyser:
Tests offered include: Chlamydia, Gonorrhoea, Trichamonas, Ureaplasma, HPV, Flu A, Flu B, RSV, SARS-CoV-2, Strep A, Strep B and Tuberculosis.
Performance Characteristics: LoD 400 copies per ml, time to result up to 35 minutes for negative but more rapid for positive (eg ten minutes)
Certification: Saudi FDA certified, Gonorrhoea/Ureaplasma (NG/UU CE certified), Chlamydia trachomatis (CE pending), Strep A (CE certified), Influenza A +Influenza B + RSV (CE certified), Trichomonas (CE certified)
Performance Characteristics:
Tests Offered | Specimen | Sensitivity | Specificity | Accuracy |
---|---|---|---|---|
CT + NG + UU | Vulvovaginal swab, Male urethral swab | 97.03% | 97.89% | 97.46% |
Flu A + Flu B + RSV | Nasal swab, Nasopharyngeal swab, Oropharyngeal swab | Flu A: 96.77% Flu B: 95.24% RSV: 100% | Flu A: 100% Flu B: 100% RSV: 100% | Flu A: 98.39% Flu B: 97.62% RSV: 100% |
GBS | Vulvovaginal swab | LoD 1000 copies/mL | ||
HPV 6/11 | Vulvovaginal swab, Male urethral swab | HPV 6: 100% HPV 11: 100% | HPV 6: 100% HPV 11: 91.3% | HPV 6: 100% HPV 11: 94.74% |
HPV 16/18 | Cervical Swab, Vulvovaginal swab | 16: 95.35% 18: 95.45% | 16: 100% 18: 100% | 16: 97.68% 18: 97.73% |
HPV 16/18/45 | Cervical swab, Vulvovaginal swab | HPV 16: 95.35% HPV 18: 95.45% HPV 45: 91.67% | HPV 16: 100% HPV 18: 100% HPV 45: 100% | 16: 97.68% 18: 97.73% 45: 97.68% |
NG + UU | Vulvovaginal swab, Male urethral swab | 97.03% | 97.89% | 97.46% |
RSV | Nasal nasopharyngeal swab, Throat swab | 97.96% | 100% | 98.31% |
SARS-CoV-2 | Nasal swab | 99% | 100% | 100% |
SARS-CoV-2 + Flu A + Flu B | Nasal nasopharyngeal swab, Throat swab | SARS-CoV-2: 96.77% Flu A: 95.24% Flu B: 100% | SARS-CoV-2: 100% Flu A: 100% Flu B: 100% | SARS-CoV-2: 98.39% Flu A: 97.62% Flu B: 100% |
Strep A | Throat swab | LoD 1000 copies/mL | ||
Tuberculosis | Tongue swab | LoD 50 CFU/mL | ||
TV | Vulvovaginal swab | 100% | 100% | 100% |